General
A 140.5$\mu$W, 99.8% Accuracy, 96.1dB SNR BioZ IC for Real-Time Cell Analysis with On-Demand Push-Pull Excitation and Continuous-Time $\Delta\Sigma$ T-D Readout
arXiv eess.SP0 views
arXiv:2610.33648v1 Announce Type: new Abstract: This paper presents a bioimpedance IC combining an on-demand push-pull driver and continuous-time $\Delta\Sigma$ T-D readout to address the trade-off between SNR, accuracy and power consumption. The driver delivers 200$\mu$App with $\leq$0.14% THD while consuming 80.3$\mu$A. Using a 6MHz clock, the readout achieves 53.6ps effective timing resolution at a 4k$\Omega$ load (0.1-4Hz noise integration). The IC achieves 96.1dB peak SNR, 99.8% magnitude accuracy and 0.13$^{\circ}$ phase error at a maximum power consumption of 140.5$\mu$W.
Source:https://arxiv.org/abs/2610.33648PreviousCross-Band Movable-Antenna Communications under Frequency-Dependent MultipathNextFixed pilot sequences can overstate the gains of learned MIMO-OFDM receivers under power amplifier distortion